Summary of Contents for ADC Adtemp 427 Temple Touch
Page 1
Adtemp TempleTouch Thermometer USE, CARE & MAINTENANCE...
Page 2
Intended Use: A Temple Touch thermometer is a device which en- ables accurate body temperature measurements ob- tained by using a unique probe, which measures the heat flow from the blood vessels to the skin and con- verts it to body temperature from the temple area in ap- proximately 6-8 seconds.
Page 3
Quick Start: 1. Press and release the MULTI-FUNC- 3. Hold firmly to temple, wait for beep. TION button. 4. Read results. 2. Wait 2 seconds. 5. Before re-measuring: wait at least 1 minute. Measuring Guidelines: 1. Measurement body location is the temple, in proximity to the temporal artery, between the end of the eyebrow and hairline.
Page 4
4. Place the thermometer's probe in the center of the temple area between the end of the eyebrow and the hairline. Gently press and hold it against the temple to ensure good contact with the skin is maintained (Figures 4a + 4b). Correct placement and good contact with the skin ensure reliable readings.
Page 5
Changing the battery: To install/replace, first switch off the device. Remove the battery cover by sliding it out of its housing. Firmly attach the batteries to their place, making sure polarity is correct. (see engraving on the bottom of battery compartment). Slide the battery cover back into place.
Page 6
Er.2: If the temperature did not register because of poor contact, patient movement or poor placement, the device will beep 5 times and display “Er.2” Remedy: Turn the unit OFF, wait 1 minute, turn it ON again and retake tempera- ture using proper technique.
Page 7
Specifications Measurement time 6-8 seconds Feature Performance / Data: Temple area of the forehead Measurement location: 95°F to 107.6°F (35°C to 42°C) Measurement range: Operation environment: 60.8°F-104°F (16° C - 40° C) Humidity: 15-95% (non condensing) Storage environment: -4°F to 122°F (-20° C - 50° C) Humidity: 15-95% (non condensing) Celsius or Fahrenheit Measurement readout:...
Page 8
Parts and Assembly Non-invasive sensor Display Window Multi-Function Button WARRANTY This Thermometer is guaranteed for a period of 24 months from date of purchase. This warranty does not cover batteries, or damage resulting from improper handling or use. Specific legal rights stipulated in your state may guide this warranty. If you have a legitimate claim under this warranty, send item(s) postage paid to Service Center, Attn: Repair Dept., 55 Commerce Dr., Hauppauge, NY 11788.
Page 9
Grand Bond Electronics Ltd Room 2208 22/F CCT Telecom Bldg. 11-19 Wo Shing Street Distributed by: ADC ® Fotan, New Territories, Shatin 55 Commerce Drive Hong Kong Hauppauge NY 11788...
Need help?
Do you have a question about the Adtemp 427 Temple Touch and is the answer not in the manual?