Selecting Output Modes; Output Controls; Example; Reading The Current Settings - HP 8112A Operating, Programming And Servicing Manual

50 mhz programmable pulse generator
Table of Contents

Advertisement

Selecting Output
Modes
Controls
Output

Example

Reading the Current
Settings
Note
DTY 50 %,WID 100 US,LEE 10.0 NS,TRE 10.0 NS,HIL 0.30 V,LOL -0.70 V
Switch off output limits
Switch on output limits
Switch off complementary output
Complement
Enable output
Disable output
OUTPUT 712*"Ll DO"
The Current Setting interrogate command
to send a response that contains all necessary commands to set the
device to its present state. The response message, called (ASCII)
learn string, can be retransmitted as a progra,m message without
alteration.
The learn string may be altered in the computer before
retransmitting it. Since the learn string is made up of the several
independent commands, it is possible to send parts of the learn string
(complete commands) to the instrument.
The description of the example program LRN-DEMO in Chapter 7
Programming Examples contains more details about the usage of the
CST command. Also refer to the example program INTERROG.
When the HP 8112A replies with a string containing all current
settings. The data is always in the same order:
Action
output
Switch on output limits and enable the
output signal.
001 #,PER 1.00 MS,DBL
Mnemonic
LO
C O
C l
DO
CST tells the HP 8112A
200 US,DEL 65.0 NS,
Programming 6-11