Procedure; Hp-Ib Verification Test; Test Setup; Equipment - HP 8112A Operating, Programming And Servicing Manual

50 mhz programmable pulse generator
Table of Contents

Advertisement

HP-IB Verification
Test

Test Setup

System
Controller

Equipment

Procedure

Note
8-30 Testing Performance
Alter one or more other parameters and note the values.
n
Check that the recalled parameter set matches the standard
n
parameter set (RCLO) except for the change made to the standard
set in step 4.
Figure 8-22. HP-IB Verification Test
Controller (HP Series 200/300)
n
Oscilloscope (HP 541211')
n
Cable Assembly BNC (2 x)
n
HP-IB cable
n
All program statements assume that the HP 8112A is at HP-IB
address 12 and that BASIC 5.0/5.1 is being used.
1. Connect the equipment as shown in the setup figure.
2. Use the following program statements to read the HP 8112A
Standard Parameter Set:
DIM A$[1611
REMOTE 712
CLEAR 712
OUTPUT 712*"CST"
ENTER 712;;$
PRINT A$
(HP 54121T)
20 dB attenuatw--
Allocate controller memory to receive HP
Set HP 8112A to remote mode
Clear HP 8112A status and select stan-
dard parameter set
Request current settings from HP 8112A
Read the HP 8112A settings
Display the HP 8112A settings