Standard Parameter Set - HP 8112A Operating, Programming And Servicing Manual

50 mhz programmable pulse generator
Table of Contents

Advertisement

Standard Parameter
Set
4-4 Operating
The Standard Parameter Set exists for two reasons:
If the instrument RAM becomes corrupted due to battery failure,
n
the Standard Parameter Set will be selected when the instrument is
switched on to give an error free display.
n
If an invalid combination of Operating and Control modes is
selected, switching the instrument off and on again, or selecting
RCL 0, will revert to the Standard Parameter Set. The Standard
Parameter Set is detailed below.
Trigger mode
Control mode
Burst
Period
Delay
Pulse width
Double pulse
Duty cycle
Leading edge
Trailing edge
High output level
Low output level
External input
Transition
Limit
Complement
Output Disable
NORM, PER, WID, DEL, illuminated (active)
NORM
Normal
CTRL
Off
BUR
0001 #
1 .OO ms
PER
65 ns
DEL
WID
100 /.Ls
200 ps
DBL
50%
DTY
LEE
10 ns
TRE
10 ns
HIL
1.00 v
0.00 v
Positive slope
EXT INPUT
Linear
LIMIT
Off
COMPL
Off
DISABLE
On