Dräger Primus Instructions For Use Manual page 73

Anaesthetic workstation software 2.n
Hide thumbs Also See for Primus:
Table of Contents

Advertisement

Leak test
Do not perform if a patient is
connected to the workstation!
When the soda lime has been
changed or
when the breathing hoses have been
changed.
Ifthe Vapor has been changed or
topped up.
e
Test with the vaporiser handwheel
set to more than 0.2 Vol.%.
@
Press the soft key »leak test« in the
standby made.
The following prompt is displayed:
»Before starting leak test, seal Y-piece
and connect sample line. Press rotary
knob to start leak test.«
Primus performs the leakage test for
volume mode/pressure mode in about
30 seconds, then system compliance is
determined for volume correction and
the overall system is checked for leaks in
the breathing system.
The breathing bag and its hose are also
tested at the same time.
Leakage is tested in the automatic
(mechanical) ventilation line (leakage,
system) and in the overall system
(leakage Man./Spont.).
Display:
The clock symbol disappears when the
test is complete and Primus displays the
values for system compliance, leakage
(system) and leakage Man. Spont. if
applicable (values >150 mL/min), see
"Leakage", page 40.
The results of the leak test are displayed
on the data screen at all times.
To return to the standby screen:
@
Press the soft key »exit«.
A new screen is displayed.
@
Close the vaporiser unit: turn the
handwheel to »0«.
Instructions for Use Primus SW 2.n
Changing patients
14 Jun.
2003
FN
Leak Test
|
15:05 160:
Leak Test | . | > si
73