Ventilation
Ventilating children
For tidal volumes VT of less than 200 mL:
@
Use paediatric hoses.
Connecting the breathing hoses
1
Use a Y-piece with connection for sample line.
@
The inspiratory and expiratory microbial filter 654 St should
not be used — this reduces system compliance.
1
—T
2
Connect a 0.5 L breathing bag with socket to the breathing
hose with the large connection sleeves. Slip the breathing
hose over the angled socket. Hang the 0.5 L breathing bag
onto the hook.
3
Slip the breathing hoses with the large sleeves onto the
inspiratory and expiratory sockets and connect the small
sleeves to the Y-piece.
4
Connect the sample line to the Y-piece and water trap.
To determine system compliance and leakage:
5
Firmly connect the Y-piece to the cone.
©
To determine system compliance and leakage,
| | )
see "Leak test" on page 73.