Dräger Primus Instructions For Use Manual page 47

Anaesthetic workstation software 2.n
Hide thumbs Also See for Primus:
Table of Contents

Advertisement

3
Press the soft key »weight«.
@
Set and confirm the weight using the
rotary knob.
Setting fresh-gas
concentrations
-
O2 concentration >O2 %<
Fresh-gas flow »flow L/min«
The fresh-gas settings can be changed
before selecting a ventilation mode.
Fresh gas does not flow in the standby
mode (soft keys = grey). The fresh-gas
flow is not enabled until a ventilation
mode has been started (soft key =
green).
Operation
14 Jun, 2003
EZ
15:05
100:
Standby
Primus
The device is tested and ready for
FRESH GAS
02
flow
%
Umin
100 | 2.00
8
Adjustment ranges and default
settings upon delivery
Fresh-gas
Adjustment range
Default setting
parameters
upon delivery
Carrier gas
| Air or N20
Air
02%
21 to 100 for carrier gas
100
Air
25 to 100 for carrier gas
N20
Fresh-gas flow| 0.2 to 18
2
L/min
Instructions for Use Primus SW 2.n
47