Dräger Primus Instructions For Use Manual page 37

Anaesthetic workstation software 2.n
Hide thumbs Also See for Primus:
Table of Contents

Advertisement

Getting started
Breathing system
@
Complete and engaged, hoses firmly plugged in.
@
Insert optional filters.
@
Fresh soda lime, without violet discoloration.
Drain any water which may have collected in the ventilator
diaphragm.
Correct operation of the workstation will be impaired by
=
condensation flowing back into the breathing system and
ventilator diaphragm.
@
install water traps in the hoses.
Care must be taken when connecting the patient.
|
|
EN
Risk of strangulation!
AGS anaesthetic gas scavenging system
1
The transfer hose from the waste gas port must be
connected.
2
The scavenging hose must be connected, the anaesthetic
waste gas probe must be plugged into the Drager wall
socket and the indicator must be green.
3
The float must be between the two marks.
instructions for Use Primus SW 2.n
37