Dräger Primus Instructions For Use Manual page 38

Anaesthetic workstation software 2.n
Hide thumbs Also See for Primus:
Table of Contents

Advertisement

Getting started
WaterLock water trap
@
Check filling level in water trap.
When the level reaches the mark:
@
Pull the water trap out of its holder and empty it.
@
Insert an empty syringe without canula, at least 20 mL, into
the connector.
Extract water, remove syringe and dispose of full syringe
with ordinary domestic waste.
e
Slide the water trap back into its holder — until it engages
tangibly.
The water trap must not be used in combination with
a medication nebulizer!
See page 147 for additional information on using the water
trap.
Preparing Primus for the self test
1
Close the Y-piece = plug firmly onto the cone.
2
Ensure that the sample line is connected to the Y-piece and
to the water trap.
3
Set APL valve to position »MÁN« and to 30 mbar.
38
Instructions for Use Primus SW 2.n