Section 2—442
Service
OPERATING
INSTRUCTIONS
INTRODUCTION
INSTALLATION
The
442
is shipped
in its carton
with the standard
accessories listed on the Accessories tab page at the end
of this manual.
At installation time,
save the shipping
carton and packaging materials for repackaging. Refer to
the end of this section for repackaging information.
Safety Considerations
This instrument may be damaged if operated with
the Line Voltage Selector switch or the Regulating
Range Selector switch set for the wrong applied line
voltage or if the wrong line fuse is used.
Refer to the Safety Summary in the front of this manual
for
power
source,
grounding,
and
other
safety
con-
siderations
pertaining to use of the instrument.
Before
connecting the instrument to a power source, verify that
the
Line
Voltage
Selector
switch
and
the
Regulating
Range
Selector switch are both set for the line voltage
being used, that the proper line fuse is installed, and that
the line cord plug matches the power source to be used.
The 442 is designed to be used with-a three-wire ac
power
system.
If a three-to-two wire adapter is used to
connect this instrument to a two-wire ac power system, be
sure to connect the ground lead of the adapter to earth
(ground).
Failure to complete
the ground
system
may
allow the chassis of this instrument to be elevated above
ground potential, thus posing a shock hazard.
REV A MAY 1980
Table 2-1
POWER
CORD
CONDUCTOR
IDENTIFICATION
Alternate
Green-Yellow |
Green-Yellow
Grounding (Earthing)
2-1