Method 2—Measure High Frequency with Two
Counters
In this method, you measure one pulse of a known width using your signal
and derive the frequency of your signal from the result. This method is good
for high frequency signals.
In this method, you route a pulse of known duration (T) to the Gate of a
counter. You can generate the pulse using a second counter. You also can
generate the pulse externally and connect it to a PFI or RTSI terminal. You
only need to use one counter if you generate the pulse externally.
Route the signal to measure (F1) to the Source of the counter. Configure the
counter for a single pulse-width measurement. Suppose you measure the
width of pulse T to be N periods of F1. Then the frequency of F1 is N/T.
Figure 7-13 illustrates this method. Another option would be to measure
the width of a known period instead of a known pulse.
NI 6238/6239 User Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
F1
Gate
F1
Ft
Source
Ft
Buffered Period
Measurement
Average Period of F1 =
Frequency of F1 =
Figure 7-12. Method 1b
7-12
Intervals Measured
T
T
...
T
1
2
K
1 2 ... N
1... ... N
... 1... ... N
1
2
K
N
+ N
+ ... N
1
2
K
K
K × Ft
N
+ N
+ ... N
1
2
K
1
×
Ft
ni.com

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ni 6238Ni 6239Daq m series

Table of Contents