P1.VCC
P1.0
P1.GND
P1.1
P1.GND
P0.0
P0.GND
Caution Exceeding the maximum input voltage or maximum working voltage ratings,
which are listed in the NI 6238/6239 Specifications, can damage the DAQ device and the
computer. NI is not liable for any damage resulting from such signal connections.

Logic Conventions

With NI 6238/6239 devices, logic "0" means that the Darlington output
switch is open, while logic "1" means closed. Table 6-1 summarizes the
expected behavior.
© National Instruments Corporation
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Buffer
P1.<0..3>
Isolators
P1.GND
P0.GND
Figure 6-2. NI 6239 Digital I/O Connections (DO Sink)