Frequency Division; Pulse Generation For Ets - National Instruments PXI-6238 User Manual

Isolated current input/current output devices
Table of Contents

Advertisement

Frequency Output can be routed out to any output PFI <6..9> or
RTSI <0..7> terminal. All PFI terminals are set to high-impedance at
startup.
In software, program the frequency generator as you would program one of
the counters for pulse train generation.
For information on connecting counter signals, refer to the
Terminals

Frequency Division

The counters can generate a signal with a frequency that is a fraction of an
input signal. This function is equivalent to continuous pulse train
generation.
For information on connecting counter signals, refer to the
Terminals

Pulse Generation for ETS

In this application, the counter produces a pulse on the output a specified
delay after an active edge on Gate. After each active edge on Gate, the
counter cumulatively increments the delay between the Gate and the pulse
on the output by a specified amount. Thus, the delay between the Gate and
the pulse produced successively increases.
Note ETS = Equivalent Time Sampling.
The increase in the delay value can be between 0 and 255. For instance, if
you specify the increment to be 10, the delay between the active Gate edge
and the pulse on the output will increase by 10 every time a new pulse is
generated.
Suppose you program your counter to generate pulses with a delay of 100
and pulse width of 200 each time it receives a trigger. Furthermore, suppose
you specify the delay increment to be 10. On the first trigger, your pulse
delay will be 100, on the second it will be 110, on the third it will be 120;
the process will repeat in this manner until the counter is disarmed. The
counter ignores any Gate edge that is received while the pulse triggered by
the previous Gate edge is in progress.
The waveform thus produced at the counter's output can be used to provide
timing for undersampling applications where a digitizing system can
sample repetitive waveforms that are higher in frequency than the Nyquist
© National Instruments Corporation
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
section.
section.
7-25
Chapter 7
Counters
Default Counter
Default Counter
NI 6238/6239 User Manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ni 6238Ni 6239Daq m series