Single Period Measurement

With single period measurement, the counter counts the number of rising
(or falling) edges on the Source input occurring between two active edges
of the Gate input. On the second active edge of the Gate input, the counter
stores the count in a hardware save register and ignores other edges on the
Gate and Source inputs. Software then can read the stored count.
Figure 7-8 shows an example of a single period measurement.

Buffered Period Measurement

Buffered period measurement is similar to single period measurement, but
buffered period measurement measures multiple periods.
The counter counts the number of rising (or falling) edges on the Source
input between each pair of active edges on the Gate input. At the end of
each period on the Gate signal, the counter stores the count in a hardware
save register. A DMA controller transfers the stored values to host memory.
The counter begins when it is armed. The arm usually occurs in the middle
of a period of the Gate input. So the first value stored in the hardware save
register does not reflect a full period of the Gate input. In most applications,
this first point should be discarded.
Figure 7-9 shows an example of a buffered period measurement.
NI 6238/6239 User Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
GATE
SOURCE
Counter Value
0
HW Save Register
Figure 7-8. Single Period Measurement
7-8
1
2
3
4
5
5
ni.com

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ni 6238Ni 6239Daq m series

Table of Contents