Maintenance - J.P. SELECTA AUTESTER ST DRY PV-B print 18L Manual

Sterilization autoclave class b 18l / 23l with a build-in printer
Table of Contents

Advertisement

Available languages

Available languages

MANUAL DE INSTRUCCIONES
CODIGO 80416 REV D
12/2023
Pag.: 49
11. Press
this button, to delete the chosen record.
12. Press
this button, so that users can search the record.
13. The hidden buttons (they appear when the mouse is close) are: the left and right scroll bars in the right screen and
the three icons in the upper right corner of the software: minimized, maximized, and closed.
14. The record is shown in the right screen.

8. Maintenance

A general review made by a technician is recommended every two years or every 2500 cycles.
8.1 Maintenance schedule
Daily:
- Clean the door gasket. This action can be performed by the user.
- Clean the sterilization chamber. This action can be performed by the user.
Weekly:
- Clean the chamber, trays and support. This action can be performed by the user.
- Clean the drain filter. This action can be performed by the user.
Monthly:
- Clean the water tank. This action can be performed by the user.
Yearly:
- General revision and maintenance. This action must be carried out by a specialist technician.
When required:
- Replace the door gasket. This action can be performed by the user.
8.2 Daily cleaning
Cleaning the door gasket
The door gasket and the mating surface should be wiped off clean each day with a clean, damp cloth. Do not use abrasive
cleaners on the gasket or mating surface. Use warm soapy water for keeping marks of sterilizer persistent, but ensure any
soap residues are completely removed by wiping both the gasket and the vessel again with water using a lint free damp cloth.
WARNING
Refer to qualified personnel for servicing: Never use a wire brush, steel wool, abrasive material, or chloride-
containing products to clean door and chamber assembly. "Caution hot surface. Avoid contact." ensure that
the sterilizer is cooled down fully before cleaning to avoid burns.
Cleaning liquid loads
Cleaning After Liquid Loads Biological media tends to boil at a higher rate than other liquids during venting. This causes
media to be spattered inside the chamber. Therefore, the chamber must be cleaned daily when you are sterilizing media.
Cleaning as follows:
- Allow unit to be cooled down.
- Wipe out chamber and door with a clean, damp cloth.
J
.P. SELECTA s.a.u.
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
Tel 34 937 700 877 Fax 34 937 702 362
e-mail: selecta@jpselecta.es - website: http://www.grupo-selecta.com

Advertisement

Table of Contents
loading

This manual is also suitable for:

Autester st dry pv-b print 23l40017384001739

Table of Contents