CAUTION:
THE DOOR SHOULD BE SHUT TIGHTLY BEFORE RUNNING PROGRAM TO AVOID THE DANGER.
ATTENTION:
THE LCD WILL DISPLAY AN ALARM IF THE DOOR IS NOT CLOSED FULLY, THE STERILIZER WILL
NOT RUN UNLESS THE DOOR HAS BEEN CLOSED COMPLETELY. IF THE DOOR HAS BEEN OPENED DURING THE CY-
CLE, THE STERILIZER WILL DISPLAY: DOOR IS OPEN DURING THE CYCLE! IF SO, PLEASE PRESS START KEY TO
CANCEL THE ALARM, THEN CLOSE THE DOOR COMPLETELY AND RESTART.
6.7 Starting a program
Close the door completely, and press START button to start a working cycle, the LCD will display the cycle state. The sterilizer will
heat, sterilize and drying the instruments automatically for you. The whole process will take 20-50 minutes. It depends on the
object being sterilized, the initial temperature, and the program you selected.
Process Display:
Cycle finished:
CAUTION:
DO NOT COVER OR PUT ANYTHING ON THE MACHINE TO KEEP HEAT WELL VENTING.
6.8 End of the sterilization cycle
When working cycle finished, the LCD Screen will give you a sound of alerting, and then you can open the door and take the
sterilized instruments out.
WARNING: Do not try to open the door when the pressure value is higher than 0.1 bar.
When the door of sterilizer opened, the program will return to the initial state, heat-preserving and waiting for next sterilizing cycle,
before start a new program it will be kept in a heat-preserving condition all the while.
CAUTION:
AFTER STERILIZING HAS BEEN FINISHED, PLEASE USE THE HANDLE TO TAKE THE TRAYS OUT
FROM STERILIZING CHAMBER. IT WILL BE BETTER TO STORE STERILIZED INSTRUMENT AFTER THEY HAVE BEEN
COOLED DOWN TOTALLY.
6.9 Apagado
If you finished the sterilization, please turn off the power switch. The power switch light will be off, and close the door but do not
lock it. If you are not going to use it for a long time, please unplug the power cord.
ATTENTION:
DURING THE STERILIZATION, WE SUGGEST THAT YOU USE THE INDICATOR TAPE. PUT THEM IN
THE CHAMBER IN ORDER TO ENSURE RELIABILITY OF STERILIZATION.
J
.P. SELECTA s.a.u.
CODIGO 80416 REV D
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.grupo-selecta.com
12/2023
Tel 34 937 700 877 Fax 34 937 702 362
Pag.: 45

Advertisement

Table of Contents
loading

This manual is also suitable for:

Autester st dry pv-b print 23l40017384001739

Table of Contents