8.3 Weekly cleaning
Cleaning Chamber, Trays and Rack
At least once a week, the trays and tray rack should be removed from the sterilizer chamber. The trays, tray rack and
chamber should be thoroughly cleaned to remove any deposits from the surfaces. Clean the trays, rack and chamber
(especially the bottom of the chamber) with appropriate antibiological cleaners. Wipe all residues from the surfaces with
a dampened, lint-free cloth.
WARNING
To prevent from collection of mineral deposits and corrosion of chamber components, use distilled or deion-
ized water only as specified. Clean chamber after each use if sterilizing saline solutions.
Cleaning water draining filter
Water draining filter might has been jammed by some dust because of use for a long-term, so effect of vacuum and drying
would be influenced. Some tiny impurity might be deposited on the filter after a long-term use, blocking the filter, so as to
influence the effect of the vacuuming and water discharging. The kinds of impurity come from smeary dust on the instru-
ments being sterilized or some calcification in the water. Keep cleaning of the inside chamber in order to make life-time of
filter much longer; please take the following advice for consideration:
- Use eligible distilled water;
- The instruments should be cleaned before placing in; it is good to use specified packing for the instruments with oil or
other impurity, don't forget to seal up.
- Rotate the water filter which composed by filter net tube(A) and filter holder(B) inside the chamber. Cleaning the part A
and B, ensure that there is no any dirty thing on it.( We suggest you to clean by ultrasonic cleaning machine). Then set it
back and rotate to the hold bottom of the chamber.
8.4 Limpieza mensual
Limpieza del tanque de agua
There are some impurities and some toxins had been left behind in the reservoir because distilled water stored for a long
time, so you need to drain and clean regularly. According to the picture which shows below, loose the screw by screw-driver,
and open the cover to clean inside.
ATTENTION
BE SURE THAT USE DISTILLED WATER PROPERLY IN ORDER TO EXTEND THE STERILIZER LIFE-TIME.
DO NOT RAVE ABOUT THE STERILIZER WHEN THE TANK HAS BEEN FILLED.
J
.P. SELECTA s.a.u.
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
Tel 34 937 700 877 Fax 34 937 702 362
e-mail: selecta@jpselecta.es - website: http://www.grupo-selecta.com

Advertisement

Table of Contents
loading

This manual is also suitable for:

Autester st dry pv-b print 23l40017384001739

Table of Contents