INSTRUCTION MANUAL
3.3 Connection to the mains
Locate an outlet (type 16A) near the autoclave. Use the network connection cable supplied or a similar one (section 1.5mm
The plug must be grounded.
Do not handle power cords with wet hands.
Check that the plug voltage matches with the indicated on the features label.
Connect the cable to the back side of the autoclave.
CAUTION:
- DO NOT BEND THE POWER CABLE TO AVOID DAMAGE
- DO NOT PUT ANY HEAVY STUFF ON THE POWER CORD
- DO NOT USE OTHER POWER CORD TO AVOID DAMAGE OF STERILIZER
- DO NOT TRY TO ADD THE POWER CORD TO AVOID ACCIDENT OR RISKS
3.4 Filling the water tank
Use deoinized water for the autoclave. Fill the tank through the top hole.
(3,5L for the 18L autoclave and 4L for the 23L one).
Another filling method is by means of the water inlet 3. For this, connect the provided hose
immerse the other end into a container with fresh water. Press key 2 for 3 seconds to start
the filling process. The equipment stops the process when it is completely filled.
3.5 Draining
The autoclave has two separate tanks, one for clean water and the other for used water.
Both tanks are separately emptied, through the valves on the right front side.
Connect a hose to open the valve.

4. Description and specifications

Autoclaves Class B are widely used in medical, dental, veterinary and generally to sterilize materials that may come in contact with
blood, as for instance, tattoo centres, piercing salons, hairdressers or beauty centres, etc.
Autoclaves Class B assure proper sterilization of different types of normal processes: solid, solid packed, wrapped porous and hollow.
To ensure sterilization, Autester ST DRY PV autoclaves achieve a complete air removal and a better steam penetration thanks to
an efficient system of fractioned vacuum.
AUTESTER ST DRY PV Clase B print are a series of autoclaves which have the following features:
-
Sterilization chamber and cover in stainless steel.
-
External epoxy coated case.
-
External steam generator to the sterilization chamber.
-
Operation digital control.
-
Specific programs for testing and autoclave diagnosis.
-
Large backlight LCD display.
-
Air inlet filter to the camera.
-
Water tank.
-
RS-232 output for printer and USB connection for logs downloading.
-
Build-in printer.
To guarantee sterilization of the load, the appropriate program indicated in the instructions must be used.
J
.P. SELECTA s.a.u.
CODE 80416 REV D
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.grupo-selecta.com
12/2023
Page: 34
Tel 34 937 700 877 Fax 34 937 702 362
)
2

Advertisement

Table of Contents
loading

This manual is also suitable for:

Autester st dry pv-b print 23l40017384001739

Table of Contents