J.P. SELECTA AUTESTER ST DRY PV-B print 18L Manual page 51

Sterilization autoclave class b 18l / 23l with a build-in printer
Table of Contents

Advertisement

Available languages

Available languages

MANUAL DE INSTRUCCIONES
CODIGO 80416 REV D
12/2023
Pag.: 51
8.5 Other maintenance
Changing Door Gasket
A plain screw driver without sharp head is needed.
Disconnect the sterilizer from the power supply. Ensure that the sterilizer is cool and depressurized.
1) Hold verge of the seal by one hand softly, and another hand should be inserted the screwdriver into the gap between
gasket and door, take out the seal slowly.
2) Once you take out one part of the seal, you can draw out the whole seal slowly. After taking out the seal, please check
and clean the groove of gasket, so does the gasket, please replace the gasket if there is some damage.
3) Fix the clean gasket in initial door groove. Attention: the gasket should be imbedded in the groove equably. At first,
please imbed the 4 spots equably into groove when fix the gasket, and then embed the other parts. After that, press the
gasket equably by hand.
4) Attention: the inner edge of gasket may be ectropium during embedding it in the door groove, at this time; you'd better
to tight it back to the groove by using the screwdriver carefully.
J
.P. SELECTA s.a.u.
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
Tel 34 937 700 877 Fax 34 937 702 362
e-mail: selecta@jpselecta.es - website: http://www.grupo-selecta.com

Advertisement

Table of Contents
loading

This manual is also suitable for:

Autester st dry pv-b print 23l40017384001739

Table of Contents