6.2

Connection of GN-10i to power source

Ensure that main power source is disconnected before making the connection
to power-supply.
For safety and stable system performance, read this instruction manual and other
necessary information carefully before performing maintenance work.
Only the qualified personnel for electric connections who have sufficient knowledge
on the structure and the risk accompanied to this equipment can perform the
following works. Ignoring this caution may result in serious body injury or death.
・Supply Voltage
・Power capacity
・Power consumption
(1) Use a circuit tester or others to confirm that the .supply voltage is within the scope of
the above.
(2) Plug the power-supply cable which comes out from the back of GN-10i
(3) Turn on the power breaker inside the GN-10i.
6.3

Dry air connection of IM-120

(1) Attach the connector for tubes to the dry air valve. Screw size is R1/4.
(2) Onto the connector attached as in the step (1), attach the tube for dry air supply.
Connect the other side of the tube to "AIR IN" connector on the back side of IM-120.
6.4

Utility connection of EMP-14W

Remove the back panel of EMP-14W and locate opening for utility connection on the
bottom. Connect cables of power-supply, nitrogen tube and remote cables of nitrogen gas
generation system through this opening.
16
WARNING
CAUTION
:100 VAC ± 10% (50/60Hz), Single-phase
:15 A or more
:Approx.400W (50/60 Hz )
ULVAC CRYOGENICS INCORPORATED
CAUTION