Instruction Manual
UMP-40W
Export Control Policy
When applying a refrigerator to a cryocooler for optical
sensors, the cryocooler falls under row 6.A.2.d.2 of the control
list established by The Wassenaar Arrangement, which is
equal to row 10(2) of appended table 1 of Japan's Export
Trade Control Order.
Customers must follow all related rules and regulations such
as Foreign Exchange and Foreign Trade Act and take
appropriate procedures when exporting or re-exporting our
refrigerators.
D/N.MJOM120E, 2022JA04
ULVAC CRYOGENICS INCORPORATED

Advertisement

Table of Contents
loading

Summary of Contents for Ulvac UMP-40W

  • Page 1 10(2) of appended table 1 of Japan’s Export Trade Control Order. Customers must follow all related rules and regulations such as Foreign Exchange and Foreign Trade Act and take appropriate procedures when exporting or re-exporting our refrigerators. D/N.MJOM120E, 2022JA04 ULVAC CRYOGENICS INCORPORATED...
  • Page 3 If you purchased our other products and/or optional devices with this product, read relevant instruction manuals carefully. Note that the UMP-40W is designed for the use outside of Japan and cannot be installed nor used in Japan.
  • Page 4 There is a risk of fire or explosion. Electrical hazard. High voltage can cause severe injury or death. Hot heating part present. There is a risk of burn injury. Caution low temperature. There is a risk of low temperature burn injury. ULVAC CRYOGENICS INCORPORATED...
  • Page 5: Table Of Contents

    PSA (Pressure Swing Adsorption) Nitrogen Generator (type GN-30i) ···· 5 Component Description ···································································· 6 Front view of UMP-40W (Figure 4-1) ················································ 6 Inside view of UMP-40W front door and lower panel (Figure 4-3 Electrical circuit) ··················································································· 8 UMP-40W Electrical circuit(Figure 4-3) ········································ 9 Flow Diagram ···············································································...
  • Page 6 Operation preparation ·································································· 31 Shutdown ··················································································· 32 When suspending for long time (for 1 to 2 weeks) and resuming UMP-40W ··································································· 32 Dispense Liquid Nitrogen Manually·················································· 33 Connecting liquid nitrogen supply flexible hose ······························· 33 Dispense liquid nitrogen ······························································· 33 Disconnection of flexible hose ························································...
  • Page 7: Safety Instructions

    The volume of nitrogen gas is 700 times of liquid nitrogen. Sealing liquid nitrogen in an airtight space produces high-pressure gas of 700 atm and may result in explosion. Use liquid nitrogen in the condition constantly open to the air. ULVAC CRYOGENICS INCORPORATED...
  • Page 8 High-pressure helium gas is filled in this unit. When disassembling or disposing of the unit or parts, be sure to discharge helium (no need to discharge helium for regular maintenance). In addition, never attempt to install the apparatus under the atmosphere of hydrochloric acid-based, chlorine gas-based, and other corrosive gases. ULVAC CRYOGENICS INCORPORATED...
  • Page 9: Disposal Considerations

    When the refrigerator unit may have pumped toxic or dangerous gases, you must contact your safety supervisor at the time of disposal, and follow the instructions to remove hazardous substances before disposing. We provide Safety Data Sheet (SDS) of our products upon your request. Please contact us when necessary. ULVAC CRYOGENICS INCORPORATED...
  • Page 10 This page intentionally left blank. ULVAC CRYOGENICS INCORPORATED...
  • Page 11: General Description Of The System

    Liquid nitrogen generator is hereafter referred to as “UMP-40W” in this book. Helium gas compressor (C30PMVRT) This unit is in the right side of UMP-40W, compress Helium gas to supply to the cold head to provide cryogenic temperatures. It is automatically turned on and off by the control circuit of UMP-40W.
  • Page 12: Component Description

    Dispense Port Emergency Stop Button (EMS) Model:UMP-40W Liquid Nitrogen Generator Figure 4-1 Front view of UMP-40W Front view of UMP-40W (Figure 4-1) Emergency stop button(EMS) :Press this button to shut down the control circuit to stop as dispensing liquid nitrogen or liquefaction. Do not use the emergency stop button except for emergency.
  • Page 13 圧 力調 整後 は、上 記表 にチ ェッ クして くだ さい 。 詳 細は 取扱 説明書 を参 照し てく ださい 。 gas pressure gauge Helium gas charge fitting (inside the panel) Figure 4-2 Inside view of UMP-40W front door and lower panel ULVAC CRYOGENICS INCORPORATED...
  • Page 14: Inside View Of Ump-40W Front Door And Lower Panel

    Terminal blocks (Power supply connection) Earth bar Figure 4-3 Electrical circuit Inside view of UMP-40W front door and lower panel (Figure 4-2) Liquid nitrogen dewar : Container that stores liquid nitrogen up to 100 litters. Cold head (RMS150T) :The cold head works in conjunction with the compressor to generate ultra-low temperature and produce liquid nitrogen.
  • Page 15: Ump-40W Electrical Circuit(Figure 4-3

    UMP-40W Electrical circuit(Figure 4-3) Circuit breaker :The Circuit Breaker is the protective breaker for control circuit of the UMP-40W. It is tripped when an electric leak or short-circuit occurs. Sequencer :The Sequencer is a controller used to control the operation of UMP-40W system.
  • Page 16: Flow Diagram

    Flow Diagram Figure 5-1 UMP-40W Flow Diagram ULVAC CRYOGENICS INCORPORATED...
  • Page 17: Specifications

    Relative humidity: 80% or less (Non-condensation other than LN dispense port)  UMP-40W must be installed indoors.  UMP-40W must not be used in the atmosphere with organic solvents or corrosive gases. PSA (Pressure Swing Adsorption) nitrogen gas generator Type : GN-30i Dimensions : 600[W] ×...
  • Page 18: Cooling Water (For Ump-40W)

    Cooling water (for UMP-40W) Flow rate : 300 – 900L/hour (recommended: 330L/h) Pressure : 0.02 – 0.17 MPa (recommended: 0.1MPa) Temperature : 10 – 32°C (recommended: 20°C) Water quality : Tap water equivalent * Check the water quality on a regular basis.
  • Page 19 Water flow rate lower limit Water flow rate Upper limit Acceptable operating range Allowable *See CAUTION on next page Temperature Lower limit (2) > 9L/min Water Flow Rate [L/min] Figure 6-1 Allowable range of cooling water for UMP-40W ULVAC CRYOGENICS INCORPORATED...
  • Page 20: Power Source For Gn-30I (When Using Gn-30I)

    Power source for GN-30i (when using GN-30i) Voltage : 200 VAC±10% (50/60Hz) , single-phase Power capacity : 15 A or above Power consumption : 1.2 kW (50/60Hz) Connection : Round type crimping terminal for M4 CAUTION Use D-type earth ground for safety. ULVAC CRYOGENICS INCORPORATED...
  • Page 21: Installation

    <NOTE> The area other than the back are required only at the time of maintenance and check, and do not always need to be cleared. 1095 Back UMP-40W Top side Front Figure 7-1 Maintenance space for the UMP-40W (4) Lock the wheels to fix the equipment. ULVAC CRYOGENICS INCORPORATED...
  • Page 22: Gn-30I Power Connection

    (3) Once the connection to the terminal board is made, attach the power cable to the circuit breaker and supply electricity by turning on the breaker in the electric circuit part of UMP-40W. Be sure to confirm that the power is supplied to the sequencers or other parts in the electric circuit of UMP-40W.
  • Page 23 When you operate multiple devices with one power source, be aware of the total capacity. Lack of capacity invites overcurrent (voltage drop) and may result in damaging the equipment. CAUTION Only the personnel who have sufficient knowledge and skill on electric connections or cables can perform the work to extend power cables. ULVAC CRYOGENICS INCORPORATED...
  • Page 24 The screw size of the connector is Rc3/8. Place seal tape on the screw part. 2) Connect the cooling water joint to the cooling water supply side. 3) Connect the cooling water hose to the water supply side and the UMP-40W water IN connector, and water discharge side and UMP-40W water OUT connector.
  • Page 25: Electric Wiring

    Electric Wiring Figure 8-1 Electric wiring UMP-40W ULVAC CRYOGENICS INCORPORATED...
  • Page 26: Touchscreen Operation

    This icon flashes when the system is in alarm state. ② ALARM Touch this button to turn to ALARM MENU and confirm ALARM the cause of alarm. Use this button to start or stop the UMP-40W. ― ③ RUN/STOP Normally, this button is in ON state. AUTO...
  • Page 27: Dewer Tank

    Suspended: Black and white Shows the state of the valve. ③ VALVE Open: Green Closed: Black and white ④ DEWER LN LEVEL Shows the volume of LN in the dewar by percentage ⑤ RETURN MAIN Return to the MAIN screen ULVAC CRYOGENICS INCORPORATED...
  • Page 28: Alarm Menu

    LN2 transfer destination is in alarm state. ALARM SUPPLY DESTN ALARM MAINTENANCE 9.4.1 ⑥ Maintenance interval is reached. ALARM MAINTENANCE ALARM ALARM 9.4.2 ⑦ Displays alarm history. HISTORY ALARM HISTORY ⑧ MAIN Move to MAIN screen. 9.1 MAIN screen ULVAC CRYOGENICS INCORPORATED...
  • Page 29 PLC UNIT[1] DSCN extended unit 1 [1]KV-N3AM A/D conversion ② [1] KV-N3AM ADC FLR error Please contact us. Inappropriate connection with ③ PLC UNIT[2] DSCN extended unit 2 [2]KV-N3AM A/D conversion ④ [2] KV-N3AM ADC FLR error ULVAC CRYOGENICS INCORPORATED...
  • Page 30 When N2 SUPPLY TIME OUT is red and blinking, there is an error in the LN supply signal. Refer to the instruction manual of the nitrogen gas generator for how to clear the alarm state. Figure 9-6 GN ALARM ULVAC CRYOGENICS INCORPORATED...
  • Page 31: Supply Destn Alarm

    These alarms can be cleared by pressing CONFIRM button. However, if the necessary maintenance is not performed, the same alarm will be activated in 120 hours. Refer to the corresponding instruction manual for the maintenance intervals of the devices. Figure 9-8 MAINTENANCE ALARM ULVAC CRYOGENICS INCORPORATED...
  • Page 32: Menu

    This screen provides information on types and timing of alarms that have taken place. Figure 9-9 ALARM HISTORY MENU Touch MENU in ALARM MENU to display the screen in Figure 9-10. ① ② ③ ④ ⑤ ⑥ ⑦ ⑧ Figure 9-10 MENU ULVAC CRYOGENICS INCORPORATED...
  • Page 33 Press GN2 MONITOR to switch to the screen shown in Figure 9-11. This screen provides information on nitrogen gas flow rate or pressure inside the dewar. The descriptions of the items are in Table 9-6 N2 MONITOR descriptions. ③ ① ④ ② Figure 9-11 N2 MONITOR ULVAC CRYOGENICS INCORPORATED...
  • Page 34 Touch LN2 DISCHARGE in MENU to display the screen in Figure 9-13. This menu is for the maintenance work. Press DISCHARGE START to discharge LN in the dewar. DISCHARGE STOP stops discharging LN Figure 9-13 LN2 DISCHARGE ULVAC CRYOGENICS INCORPORATED...
  • Page 35 This menu is for our service personnel to check the status when a trouble arises. Figure 9-14 PLC MENU DIO MONITOR, [1] AIO MONITOR, [2] AIO MONITOR are shown in Figure 9-15, Figure 9-16, and Figure 9-17. Figure 9-15 DIO MONITOR [1]AIO MONITOR Figure 9-16 [2]AIO MONITOR Figure 9-17 ULVAC CRYOGENICS INCORPORATED...
  • Page 36 Figure 9-19. Correct the date or time as desired and press ENTER, and then SET to complete the change. Figure 9-18 SET TIME① Figure 9-19 SET TIME ULVAC CRYOGENICS INCORPORATED...
  • Page 37: Operation

    Operation preparation Start nitrogen gas supply with GN-30i (1) Press RUN/STOP and GN ON/OFF of UMP-40W touchscreen (not of GN-30i) to start up GN-30i. When it starts operation normally, the GN ON/OFF on UMP-40W touchscreen flashes of 1 second each.
  • Page 38: Shutdown

    75L (75%), and stops when it reaches to 100L (100%). This cycle goes on repeatedly. Shutdown When the UMP-40W is turned OFF with the RUN/STOP, RUN/STOP lights off and all the operations stop including LN production, supplying nitrogen gas, or dispensing LN...
  • Page 39: Dispense Liquid Nitrogen Manually

    Connect the joint (coupler joint) of attached flexible hose to LN manual dispense port located at the front of UMP-40W. When connecting to the coupler joint, do so while pushing the outer ring part of the UMP-40W side. Dispense liquid nitrogen To manually dispense liquid nitrogen, insert the head end of flexible hose in liquid nitrogen container, and keep pressing the MANUAL START button for 3 seconds.
  • Page 40: Disconnection Of Flexible Hose

    O-ring inside the coupler joint. Disconnect flexible hoses in the opposite way of connecting. Push down the outer ring of the UMP-40W side joint and pull out the flexible hose to disconnect. ULVAC CRYOGENICS INCORPORATED...
  • Page 41: Maintenance And Inspection

    ✓ GN-30i Solenoid valve  1. Check the total operation time of UMP-40W with the hour meter on operation display.  2. The total operation time of GN-30i is shown in the sequencer display in the electric circuit of UMP-40W. The operation light of GN-30i blinks when the maintenance time approaches.
  • Page 42 Since helium is filled in this equipment as well as in the maintenance parts (cold head, oil adsorber), do not disassemble these parts. When you have to disassemble or dispose such parts, discharge helium using the optional charging adapter kit. ULVAC CRYOGENICS INCORPORATED...
  • Page 43: Troubleshooting

    Please supply the power fails to come on. The power cord is not connected. Connect the power cord correctly. The circuit breaker of UMP-40W Turn ON the circuit breaker. is turned OFF. (2) Power breaker is tripped. Short-circuit or electric leakages Please contact us.
  • Page 44 Ice or frost adheres to inside the Please contact us. dewar or around the cold head. The purity of Nitrogen gas is not Check the purity of nitrogen gas sufficient. supplied. (6) Other failures Please contact us. ULVAC CRYOGENICS INCORPORATED...
  • Page 45: Accessories

    Nitrogen Gas Tube 1 (5m) Single Head Wrench 1 set Instruction Manual 1 (This book) C30PMVRT Instruction Manual GN-30i Remote Cable for Nitrogen Generator Instruction Manual  The above accessories may vary depending on the customer’s specifications. ULVAC CRYOGENICS INCORPORATED...
  • Page 46: Warranty

    As a general rule, diagnosis of failure should be done on site by customer. However, ULVAC CRYOGENICS or our service network can perform this service for an agreed fee upon the customer’s request. There will be no charge if the cause of the breakdown is found to be a fault of ULVAC CRYOGENICS.
  • Page 47 ULVAC CRYOGENICS products and compensation for damages to products other than ULVAC CRYOGENICS products and other services are not covered under warranty. 3. Repair period after production is discontinued ULVAC CRYOGENICS shall accept product repairs for seven years after production of the product is discontinued. Manufacturer: ULVAC CRYOGENICS INCORPORATED For our contact information, refer to “SERVICE NETWORK”...
  • Page 48 This page intentionally left blank. ULVAC CRYOGENICS INCORPORATED...
  • Page 49 ULVAC CRYOGENICS INC. www.ulvac-cryo.com 1222-1 Yabata, Chigasaki, Kanagawa 253-0085, Japan <Sales> Tel: +81-467-85-8884 <Service Engineering Division> Tel: +81-467-85-9366 Fax: +81-467-83-4838 ULVAC CRYOGENICS KOREA INC. www.ulvac-cryo.co.kr 107, Hyeongoksandan-ro, Cheongbuk-Myeon, Pyeongtaek-si, Gyeonggi-Do, Korea, 17812 Tel: +82-31-683-2926 Fax: +82-31-683-2956 ULVAC CRYOGENICS (NINGBO) INC.
  • Page 50 This page intentionally left blank. CS-2 ULVAC CRYOGENICS INCORPORATED...
  • Page 51 2017 / 09 / 27 2017SR02 6.3.2 N2 ALARM Figure 6-5 and Table 6-4 have been modified. 2019 / 09 / 24 2019SR03 Full-fledged revision 2022 / 01 / 18 2022JA04 12. Maintenance and Inspection has been modified. ULVAC CRYOGENICS INCORPORATED...
  • Page 52 This page intentionally left blank. ULVAC CRYOGENICS INCORPORATED...