3.11.3 Adjusting camera diopter

Adjust camera diopter using the following steps.
TIP:
Be sure to adjust camera diopter, to prevent the following issues.
1. Parfocal among objective lenses getting worse.
2. Auto-focusing performance getting worse.
1.
Put a specimen on the stage.
2.
Set the highest magnification objective lens in the light path.
Command: 1OB
3.
Move the Z-stage and adjust focus on the specimen.
4.
Set the lowest magnification objective lens in the light path. Be sure not to move the stage height.
Command: 1OB
5.
Move the circular dovetail of the camera adapter and adjust focus on the specimen.
21
3. Setup procedures
Laser Condition
Lock (Laser disable)