Olympus BXC-CBB Instructions Manual page 24

Modular microscope assemblies hardware
Table of Contents

Advertisement

Item
Voltage Range
-10 to +10V
Data rate
5 msec
Zero cross point
1
linearity
Monotonic
Gain Adjustment 0.1 to 10X
Clipping Voltage -10 to +10V
Signal Sign
Positive: Above the focal point (Near)
Negative: Below the focal point (Far)
(3) Focus Status Judgement Signal
Label: S-LOCK
Description: In-Capture and In-Focus signals.
Function: Informing Focus status ("H" ≥ 4.0 V, "L" ≤ 0.5 V )
Connector type: D-SUB15 (female)
Pin No.
Pin Name
Pin No.
1
N.C.
2
N.C.
3
N.C.
4
N.C.
5
N.C.
6
In-Capture signal
7
N.C.
BXC-CBE1 has sets of default values of TH(In-capture) and TH(In-focus), corresponding to each objective lens. The
l
values can be changed.
In-Capture / In-Focus / GND pin assignments are the same as U-AFA2M. However, other signals are assigned to
l
different pins. Therefore, all pins except In-Capture/In-focus/GND signals must be N.C.(No Connect)
3. Setup procedures
Specifications
Pin Name
9
N.C.
10
GND
11
N.C.
12
N.C.
13
N.C.
14
In-Focus signal
15
N.C.
Comments
10V: (A-B) / (A+B) = 1
-10V: (A-B) / (A+B) = -1 #gain=1
U-AFA2M: 10 msec
Around focal point
Our standard sample
+0.4>(A-B) / (A+B)>-0.4
Adjustable via RS232C
Adjustable via RS232C
Reversible via RS232C
20
3.11 Setting up BXC-FSU

Advertisement

Table of Contents
loading

This manual is also suitable for:

Bxc-cbe1Bxc-fsu

Table of Contents