Initial Setup; Ambulance Information; Accessories; Charge And Install The Battery - Ferno iNSX Integrated Patient Transport & Loading System Instructions For Use Manual

Table of Contents

Advertisement

7.1 Ambulance Information

The ambulance bumper extension must not exceed 13" (330 mm). The
patient compartment must have a level floor large enough for the fully
lowered cot, and a Ferno® iN∫LINE® fastening system installed.

7.2 Accessories

Before placing the cot in service, assign appropriate personnel to install
any accessories shipped with the cot. Keep accessory users' manuals
with this manual for future reference. To order accessories, see page
46. For additional free manuals, contact Ferno (page 2).

7.3 Charge and Install the Battery

Immediately charge the battery upon receipt. Use only the provided
battery charger or an ICS-powered fastening system (Figure 52).
The charger is compatible with standard household electricity and is
supplied with an input cable for your country.
For complete information, see page 16, page 17, or the supplied
charger and battery manuals.

7.4 Set the Loading Height

The user can set the cot so it stops automatically at the proper loading
height for the assigned ambulance.
Be aware: The user-set loading height cannot be set when the cot is at
the factory-set maximum transport height or maximum height.
To set or change the loading height:
1.
On a level surface, raise or lower the cot until the loading wheels
roll directly onto the ambulance floor (Figure 53). Confirm the
loading wheels are resting on the ambulance floor, the safety bar
is engaged with the safety hook, and the loading-leg wheels have
not lifted off the ground.
2.
Press the Mode-Select button four times to enter the set load-
height mode (Figure 53 inset). See "Mode-Select Button" on page
18.
3.
Press and hold
to set the loading height before the 15-second
countdown timer expires. When pressed, a new 5-second timer
will appear on the display. Continue holding
timer expires.
4.
Turn the power switch OFF and then ON. See "Power Switch" on
page 14.
5.
Retract and extend the legs to verify the cot stops at the new
loading height. Test loading and unloading the cot with the
ambulance.
44
iN∫X®

7 - INITIAL SETUP

until the 5-second
WARNING
Attaching improper items to the cot can cause injury.
Use only Ferno-approved items on the cot.
NOTICE
Loose items and debris on the patient compartment floor
can interfere with locking the cot into the iN∫LINE. Keep the
patient compartment floor clear.
5
© Ferno / 234-3622-09 / December 2022
52
53

Advertisement

Table of Contents
loading