Maintenance; Maintenance Schedule; Service Life; Disinfecting /Cleaning Restraints - Ferno iNSX Integrated Patient Transport & Loading System Instructions For Use Manual

Table of Contents

Advertisement

6.1 Maintenance Schedule

The cot requires regular maintenance. Set up and follow a maintenance
schedule. The table at right represents minimum intervals for
maintenance. Keep maintenance records. For a sample record sheet, see
"Maintenance Record" on page 47.
When using maintenance products, follow the manufacturers' directions
and read the manufacturers' material safety data sheets.

6.2 Service Life

Many factors affect a product's service life, including frequency of use
(high-volume versus low-volume ambulance service); local environment
(extreme temperature, high humidity, etc.); maintenance (adherence to
the inspection and maintenance schedule); and other factors.
Follow the disinfecting, cleaning, and inspection procedures in this
manual to ensure long life for the cot.

6.3 Disinfecting /Cleaning Restraints

Remove the restraints from the cot before cleaning/disinfecting them.
To disinfect: Apply disinfectant to all restraint surfaces, following the
disinfectant manufacturer's instructions for application method and
contact time. Wipe with a clean cloth. Do not immerse metal buckles or
slide bars in liquid.
To clean metal buckles: Apply a disinfectant cleaner to the metal buckle,
following the disinfectant manufacturer's instructions for application
method and contact time. Wipe with a clean cloth. Do not immerse the
metal buckles in liquid.
To clean the webbing:
1.
Add a disinfectant cleaner to warm water, following the
disinfectant manufacturer's instructions.
2.
Immerse webbing in the solution for the time directed by the
disinfectant manufacturer. Do not immerse the metal buckles or
slide bars in the solution.
3.
Repeatedly dip the webbing in clear water to rinse. Do not
immerse the metal buckles or slide bars in liquid.
4.
Hang the restraint to air-dry.
5.
Attach only dry restraints to the equipment.

6.4 Disinfecting/Cleaning Mattress

Remove the mattress from the cot and inspect it. Replace the mattress if
it is ripped, punctured or damaged.
To disinfect: Apply disinfectant to the mattress, following the disinfectant
manufacturer's instructions for application method and contact time.
To clean: Wash the mattress with warm, soapy water and a soft cloth.
Rinse the mattress with clear water. Hang the mattress to dry, or dry it
with a towel. To attach the mattress to the cot, see page 21.
© Ferno / 234-3622-09 / December 2022
iN∫X®

6 - MAINTENANCE

Improper maintenance can cause injury and damage.
Maintain the cot and battery only as described in this
manual.
Minimum
Maintenance
Daily Operator Duties (page 8)
Disinfecting (pages 35-36)
Cleaning (pages 35-36)
Pressure-Washing (page 36)
Inspecting (page 37)
Manual Back-up Cable (page 33)
Backrest Cable (page 34)
Disinfectants and cleaners containing phenolics or iodines
can cause damage. Do not use products containing these
chemicals. If bleach is used, rinse with plenty of clear water to
remove bleach residue.
Using abrasive cleaning compounds or applicators on the cot
can cause damage. Do not use abrasive materials to clean the
cot.
Metal buckles and slide bars can be damaged by immersion
in liquids or by washing-machine action. Disinfect and clean
only as described in this manual.
Water under high pressure, or steam, can penetrate joints,
flush away lubricant, and cause corrosion. Use caution when
cleaning moving parts such as joints and hinges.
WARNING
Each
Use or
As
Daily
Needed Monthly
NOTICE
35

Advertisement

Table of Contents
loading

Table of Contents