4.6 Battery

The iN∫X battery (J, Figure 6) is a rechargeable, 36 Volt DC lithium-iron
phosphate battery (LiFePO4). The battery has a carry handle (K). The
battery installs under the shock frame, behind the display (Figure 7).
For complete use, maintenance, and safety information, refer to Ferno
publication 234-3595.
Charge the battery by securing the cot in a fastening system with
powered ICS, or remove the battery from the cot and connect it
directly to a Ferno battery charger.
Charging time is 15 minutes (one transport, maintaining the charge
with ICS) to 1 hour, 45 minutes (depleted battery). If a battery has a
very low charge, extend the charging time.
Actual charging time varies based on frequency of recharge,
depth of discharge, and battery age. Patient weight and extreme
temperature can also affect the power available from each battery
charge.
Turn the power OFF if the cot will not be used for more than 30
minutes.
The battery is water-resistant and is suitable for outdoor use. Do
not submerge the battery in water.
Before placing the battery in storage, charge it. Store only a fully-
charged battery. See page 39.
Replace a damaged or cracked battery. Do not open the sealed
battery case. Do not dismantle, open or shred the battery.
Recycle the battery when it no longer holds a charge (page 39).
REMOVING THE BATTERY
1.
Turn the power switch OFF (A, Figure 3, page 14).
2.
Raise the shock frame.
3.
Place your thumbs under the front corners of the battery cover
and lift to remove.
4.
Grasp only the battery-cable plug (L, Figure 7) to disconnect the
cable from the battery. Do not pull on the cable.
5.
Use the battery carry handle to lift the battery out of the cradle.
ATTACHING THE BATTERY
6.
Confirm the power switch is OFF.
7.
Raise the shock frame.
8.
Place your thumbs under the front corners of the battery cover
and lift to remove.
9.
Angle and slide the battery into the cradle. Press down to lock the
battery in place in the cradle.
10. Connect the battery cable (L, Figure 7) to the battery port.
11. Insert the front tabs of the battery cover into the slots and then
press down the area near the display to snap the cover into place.
12. Lower the shock frame.
16
iN∫X®
J
K
If the battery sustains noticeable or suspected damage or
is hot to the touch, immediately remove it from service and
store outside of a building.
NOTICE
L
© Ferno / 234-3622-09 / December 2022