4.4 Display Zones

Turning the power switch ON ( ) (A, Figure 3 on page 14) illuminates
the display and prepares the cot for its default, powered operating
mode. The display is off when the power switch is OFF (0)or when the
cot enters Sleep Mode. To awaken from Sleep Mode, press . See "Sleep
Mode" on page 18.
The display shows the following information (Figure 5):
COT: The cot is displayed in various colors to indicate the current
status of each area. Refer to the table at right.
TIMER: The countdown timer appears in the lower-left corner of the
display only when the cot is in a direct power mode. See "Direct
Power Modes" on page 31. If the countdown timer is allowed to
expire, the cot returns to its default powered mode.
STATUS: This bottom center of the display illuminates with notices
and safety information. Based on the condition, a warning triangle,
other symbol, and/or alphanumeric code may be displayed. The
warning triangle is yellow for medium-level notices, and red for
high-priority notices (Figure 5). See "Audible Alarm" below and
"Status Indicators" on page 45.
BATTERY: The symbol indicates the current battery charge. It is
illuminated whenever the cot power is ON or the cot is connected
to a powered ICS. The color and shape of the battery indicator
shows the approximate charge remaining in the battery. Refer to
the table at right.
The cot beeps when a status indicator is displayed. The
severity of the status determines the number of beeps.

4.5 Audible Alarm

The cot will sound a series of beeps when a status indicator is displayed.
The number of beeps is dependent on the severity of the situation. The
alarm continues to sound until the status returns to normal.
High-priority: 2+3 beeps, pause, repeat
Medium-priority: 2 beeps/15 seconds
Low-priority: 1 beep/minute
© Ferno / 234-3622-09 / December 2022
NOTICE
iN∫X®
Cot Color Indicators
LEG
Green
White
Blue
PATIENT SURFACE
White
Yellow/Red
WHEELS
Green/White
Red
N04
1
2
Symbol
Definition
1
Connected to ICS, charging
2
Connected to ICS, charge full
3
Full charge
4
Moderate charge
5
Low charge
6
Battery Depleted (flash red/outline)
Legs may be extended or retracted
Legs will not extend or retract
Direct Power: blue legs move when a
button is pressed
Safe transport height
Above max transport height. Lower
before rolling.
Cot may be rolled if patient surface is at
a suitable transport height
Do not roll the cot
3
4
5
6
5