ADJUSTING THE HOLDER
1.
Use a 3 mm hex wrench to remove the four mount screws from
beneath each mount (XX, Figure 50). Retain the mount screws.
2.
Align the mounts as needed based on the size of the oxygen
cylinder bottle (Figure 51).
3.
Lay the cylinder fastening straps across the assembly, with the
D-rings facing the cot control end and with the strap between the
mounting holes for the selected bumper mount location (Figure
50 inset).
4.
Prepare the 8 mount screws (XX) and receiving threads in the
mounts:
Clean the threads of the screws, plus the receiving threads in the
mounts with acetone or an appropriate threadlocker cleaner/
primer (not supplied). Allow the threads to dry completely.
Apply 1 drop of removable threadlocker (not supplied) to the
threads of each screw. Install the screws immediately. Wipe off
any excess threadlocker.
5.
Use a 3 mm hex wrench to install and tighten four mount screws
into each mount (Figure 50).
INSTALLING THE HOLDER
1.
Prepare the 4 short screws and receiving threads in the hex
spacers (TT, Figure 49 on page 42):
Clean the threads of the screws, plus the receiving threads in the
hex spacers (YY) with acetone or an appropriate threadlocker
cleaner/primer (not supplied). Allow the threads to dry
completely.
Apply 1 drop of removable threadlocker (not supplied) to the
threads of each screw. Install the screws immediately. Wipe off
any excess threadlocker.
2.
Place a rubber washer (WW) on each hex spacer (YY, Figure 49,
page 42).
3.
Insert the oxygen cylinder mount (SS) over the four hex spacers.
Confirm the mount is properly seated on the hex spacers.
4.
Place one small washer (UU) and one large washer (VV) on each
short screw and insert the screws through the oxygen cylinder
mount and into the hex spacers (Figure 49 on page 42).
5.
Use a 4 mm hex wrench to tighten the short screws.
6.
Confirm the oxygen cylinder mount holds the cylinder securely.
See page 23.
© Ferno / 234-3622-09 / December 2022
iN∫X®
Allow threadlocker to dry for a minimum of one hour before
using the cot. Full cure time is 24 hours.