If You Encounter a Problem

If you have difficulty when installing or using the analyzer, check the following list of
commonly encountered problems and troubleshooting procedures. If the problem that you
encounter is not in the following list, refer to additional troubleshooting sections in the
Service Guide.

Power-Up Problems

If the analyzer display does not light:
• Check that the power cord is fully seated in both the main power receptacle and the
analyzer power module.
• Check that the AC line voltage selector switch is in the appropriate position
(230 V/115 V) for your available power supply.
• Check that the analyzer AC line fuse is not open.
WARNING
For continued protection against fire hazard, replace the fuse with
the same type and rating.
Refer to
Figure 2-17
continuity light or an ohmmeter to check the fuse. An ohmmeter should read very close
to zero ohms if the fuse is good. For 115V operation, use a 5A 125V TD fuse (HP part
number 2110-1059). For 230V operation, use a T 4A 250V fuse (HP part number
2110-0734).
• Contact the nearest Hewlett-Packard office for service, if necessary. A list of
Hewlett-Packard sales and service offices is provided in
Figure 2-17 Line Fuse Removal and Replacement
Chapter 2
to remove the fuse from the power module. You can use a
Quick Start: Learning How to Make Measurements
If You Encounter a Problem
Table 2-1 on page
2-26.