HMP80 Series User Guide
10. Technical data
10.1 HMP80 specifications
Table 6 HMP80 series measurement performance
Property
Relative humidity
Measurement range
Accuracy at +23 °C (73.4 °F)
Factory calibration uncertainty
T
response time
63
Sensor
Temperature
Measurement range
Accuracy at +23 °C (+73.4 °F)
Factory calibration uncertainty
Sensor
1)
Defined against calibration reference. Including non-linearity, hysteresis, and repeatability.
2)
Defined as ±2 standard deviation limits. Small variations possible; see calibration certificate.
3)
Exposing temperature sensor to temperatures below −20 °C (−4 °F) may cause permanent
additional deviation of ±0.1 °C (0.18 °F).
42
Description/Value
0 ... 100 %RH
±0.8 %RH (0 ... 90 %RH)
1)
±0.5 %RH (0 ... 40 %RH)
2)
±0.8 %RH (40 ... 95 %RH)
15 s
HUMICAPâ R2C
HMP80N: −20 ... +60 °C (−4 ... +140 °F)
HMP80L: −50 ... +120 °C (−58 ... +248 °F), short-
time measurement range
−50 ... +180 °C (−58 ... +356 °F)
±0.1 °C (±0.18 °F)
1) 3)
±0.1 °C (±0.18 °F) at +23 °C (+73.4 °F)
2)
Pt100 RTD Class F0.1 IEC 60751
M212903EN-A