6.3 Configuration options
The complete probe configuration is accessible via the Insight software. After connecting the
probe, select
> Configure device to access the configuration options.
If Insight has been set to Advanced mode, additional configuration options may be available.
Use Advanced mode only when instructed to do so by product documentation or Vaisala
technical support.
Figure 9 HMP80L in Insight software
More information
‣ Restoring factory default settings with Insight software (page 40)
6.4 Diagnostics in Insight
Messages sent by the connected devices are shown automatically in Insight. Additional
diagnostic data is available on the Diagnostics page. The data available depends on the probe
model and Insight operating mode (basic or advanced). The data can be very useful when
diagnosing issues together with Vaisala support, particularly the diagnostic files.
Diagnostic files
If Insight is in Advanced mode, you can retrieve the following diagnostic files from the
connected probe:
• SSR/T
histogram: Table of humidity and temperature conditions measured by the probe
s
• Error log: Cumulative total of the error and status events tracked by the probe
Both files contain data from since the probe was manufactured, or since the files were last
cleared. Clearing the files is not recommended unless instructed by Vaisala support. The
buttons to clear the files are located at the bottom of the Diagnostics page.
Chapter 6 – Configuration with Insight software
27