HMP80 Series User Guide
9. Troubleshooting
9.1 Problem situations
Table 5 Troubleshooting table
Problem
Measurement output seems
incorrect
Values of measurement
parameters stop changing for a
few minutes
9.2 Error messages
Probes can communicate with error messages when they are connected to Indigo80 indicator
or Insight software. The messages are categorized according to the severity of the status:
• Critical errors are fatal to the operation of the device. It may not be able to respond to
communication at all, and will not measure correctly.
• Errors prevent normal operation of the device. Depending on the problem, errors may
resolve themselves. For example, a completely wet humidity sensor may cause a humidity
measurement error.
• Warnings do not prevent normal operation but may indicate possible problems.
• Status indicates a known state of the device.
38
Possible cause
Installation location is not
representative of actual
conditions you want to
measure
Heat conduction along probe
shaft is interfering with
measurement accuracy
Probe is in need of adjustment
Probe is performing a sensor
heating function such as sensor
purge or waiting for the sensor
to cool down
M212903EN-A
Solution
Verify the installation location
and relocate the probe if
necessary.
Follow the installation
recommendations for cases
when temperature of measured
environment differs greatly
from ambient. See
Measurement location
(page
21).
Calibrate and adjust the probe.
Wait for the function to
complete and measurement
parameters to be available
again.