Warnings
Calibration has expired [353]
Software restart [679]
Unexpected device restart
[216]
Status messages
Calibration is about to expire
[352]
Purge in progress [287]
The readings stay frozen until
the sensor cools down [288]
More information
‣ Technical support (page 51)
9.3 Restoring factory default settings with Insight
software
If needed, you can restore the probe back to its factory default settings using the Insight
software. Doing this will also clear any user adjustment and restore the latest adjustment
performed by Vaisala.
1. Connect the probe to the Insight software.
2. Select
> Factory default settings > Restore settings.
3. Wait for the probe to be re-detected.
More information
‣ Configuration options (page 27)
40
This warning is displayed by
the calibration reminder
functionality if the calibration
interval has been exceeded.
The probe has automatically
restarted itself.
This message is displayed by
the calibration reminder
functionality.
Sensor purge of the humidity
sensor is ongoing.
Probe is waiting for its sensors
to cool down.
M212903EN-A
Calibrate the probe and update
the calibration date
information.
Check that supply voltage is
stable and operating
environment is within
specification.
Calibrate the probe and update
the calibration date
information.
Wait for sensor purge to
complete.
Wait for measurement readings
to become available.