HMP80 Series User Guide
7.3.2 Performing manual adjustment with Indigo80 handheld
indicator
• Probe connection cable
• Reference environment(s) for producing the desired reference values
This procedure tells you how to use the indicator to adjust the probe's measurement in 1 ... 2
points.
1. Connect the probe(s) to the indicator.
2. Open the main menu by pressing
3. Select Devices. If you have more than one device connected to the indicator, make a
further selection between the devices.
4. If your probe supports the sensor purge feature, wait for the start-up purge to finish or
start the purge manually by selecting Devices > <your device> > Sensor purge.
After the purge has been completed, continue to the next step.
5. Select Calibrate / Adjust.
6. Select the parameter, for example, Relative humidity.
7. Select Manual adjustment.
8. Place the probe to the first reference point and select Reference value, point 1 to open
the stabilizing graph view. Wait for the measurement to stabilize fully.
The stabilization typically takes at least 10 ... 15 minutes.
a. When the reading has stabilized, press
b. Use the arrow buttons to set the value of the reference, and then select OK or Set.
If you already have the reference and measured values available for your
calibration point(s), you can just skip the stabilizing graph view with the
button to get to the view where you can set the reference value. Then replace
the automatically filled Measured value, point 1 and Measured temperature,
point 1 with your actual values.
9. If you want to adjust the measurement in 2 points, place the probe to the second
reference point. Then scroll down the <parameter> adjustment view and select
Reference value, point 2 to open the stabilizing graph view. Wait for the measurement to
stabilize fully.
a. When the reading has stabilized, press
b. Use the arrow buttons to set the value of the second reference, and then select OK or
Set.
32
.
.
.
M212903EN-A