Accessing certain configuration options for your probe requires using the free
Insight PC software, downloadable at www.vaisala.com/insight.
1. Connect the probe(s) to the indicator.
2. Open the main menu by pressing
3. Select Devices. If you have more than one device connected to the indicator, make a
further selection between the devices.
4. Select Settings to access and change the features available for your probe.
5. Exit the menu by pressing
5.4.1 Example of configuring filtering factor with Indigo80
handheld indicator
• Probe connection cable
The filtering factor allows averaging the output if the measuring environment produces
occasional exceptionally high or low readings.
This example shows how to configure the filtering factor as 0.9 using the Indigo80 handheld
indicator.
This means that the new measurement output is a combination of the latest measurement
(90 %) and the previous measurement (10 %).
1. Connect the probe(s) to the indicator.
2. Open the main menu by pressing
3. Select Devices. If you have more than one device connected to the indicator, make a
further selection between the devices.
4. Select Settings > Filtering.
5. Select Filtering factor.
a. Use the arrow buttons to set the value as 0.9.
b. Select Set.
6. Select Extended filtering on/off > On to enable measurement filtering using the filtering
factor defined in the previous step.
7. Exit the menu by pressing