10. In <parameter> adjustment view, check the difference between each reference and
measured value. Very large differences may be due to insufficient stabilization time or
unsuitable calibration setup.
• To store the adjustments to the probe, scroll down and select Activate adjustment >
OK. Confirm by selecting OK.
• To exit without taking the adjustment in use, press
11. Update the calibration information in Calibrate / Adjust > Calibration information.
Navigate to the Calibration information menu by pressing
12. Press
to end calibration.
7.3.3 Restoring factory adjustment with Indigo80 handheld
indicator
• Probe connection cable
If needed, you can restore the factory adjustment for your probe using the Indigo80 handheld
indicator.
Note that restoring the factory adjustment is not necessary before a new adjustment: you can
just perform the new adjustment on top of the old adjustment. However, you can select to do
the restore procedure if your adjustment fails for some reason, or if you want to return to the
original factory adjustment.
1. Connect the probe(s) to the indicator.
2. Open the main menu by pressing
3. Select Devices. If you have more than one device connected to the indicator, make a
further selection between the devices.
4. Select Calibrate / Adjust.
5. Select any of the available parameters to access the view where you can restore the
factory adjustment.
6. Scroll down and select Restore factory adjustment > OK.
7. Confirm by selecting OK.
8. Press
to end calibration.
Chapter 7 – Calibration and adjustment
.
to end calibration.
.