HMP80 Series User Guide
5.5 Configuring environment settings with
Indigo80 handheld indicator
• Probe connection cable
For optimal measurement accuracy, you can use the Indigo80 handheld indicator to give
additional information about the connected probe's measurement environment.
The environment settings available via Indigo80 correspond to the environmental
compensation settings of the probe. Therefore, Environment settings and
Settings > Compensation setpoints are interconnected: the configuration set in
either menu is applied to both.
1. Connect the probe(s) to the indicator.
2. Open the main menu by pressing
3. Select Devices. If you have more than one device connected to the indicator, make a
further selection between the devices.
4. Select Environment settings to access the settings for your probe.
The available environment settings depend on the probe and probe firmware version.
Pressure
Value for pressure compensation. The entered value is persistent and remains
unchanged at device restart.
Temperature
If any value is entered in this field, the device will use it as its temperature
measurement value instead of its own sensor reading. This is needed when sensor
warming is used, as it will also heat the temperature sensor of the device.
At device restart, the value is cleared.
24
.
The temperature value needs an external source, for example, another
probe that is able to provide the value constantly in real time, and the
value needs to correspond to the temperature of the measurement
environment as accurately as possible. Thus, set the temperature value
only if you have another probe connected to the indicator from which
you can read the value. Do not set the temperature value manually.
M212903EN-A