5.2 EXIT FROM ERROR DISPLAY

By pressing the STOP key, you exit from error display and go back to
stand-by mode.
If error is still displayed, turn off the main switch of the centrifuge and
turn it on again.
If the centrifuge still doesn't return to stand-by mode, call service!

5.3 USER'S PARAMETERS

With user's parameters, you can check and set some of the operational properties of
the centrifuge. For entering to user's parameters, the centrifuge must be in stand-by
mode. Simultaneously press the LEFT and RIGHT keys. The first parameter shows on
display (ror).
Then you can select the parameters by pressing the LEFT and RIGHT keys. If you want
to change the setting of the parameter, press the PROGRAM key, to display the current
value of the parameter. Then change the setting of the parameter by pressing the UP
and DOWN keys and confirm it by pressing on the PROGRAM key. The name of param-
eter, which you've just set,is shown again. Then you can continue the setting of user's
parameters using the above procedure, or you can exit to stand-by mode by pressing
the START key.
DISPLAY
PARAMETER SETTING
ror
rSP
SLI
OLI
bSE
tIn
bEP
Radius (mm) of currently used rotor. The value can be
86
set by pressing the UP and DOWN keys. (25 - 200 mm)
Maximal permitted speed for currently used rotor.
15500
(6000 RPM or 15500 RPM)
Automatic start of operation at lid closing. The setting
dIS
can be chosen by pressing the UP and DOWN keys.
(dIS = disable, Enb = enable)
Automatic opening of the lid at operation stop. The
dIS
setting can be chosen by pressing the UP and DOWN
keys. (dIS = disable, Enb = enable)
Block the changing of operation parameters during
the operation of the centrifuge. The setting can be
dIS
chosen by pressing the UP and DOWN keys. (dIS =
disable, Enb = enable)
dEC
Decreasing or increasing of time display. The setting
can be chosen by pressing the UP and DOWN keys.
(dEC = decreasing, InC = increasing)
Enb
Turn on the beeper. The setting can be chosen by
pressing the UP and DOWN keys. (dIS = disable, Enb =
enable)