Table of Contents

Advertisement

After installation, it is recommended that you wait for some time,
before connecting the centrifuge to the mains power supply. This pre-
vents damage to electronic components due to condensation, which
can occur, when you bring the device from a cold environment to a
warm environment.
Before connecting the centrifuge to the mains power supply, check that the mains
voltage and frequency correspond to the specifications on the name label of the
centrifuge. The mains cable of the centrifuge may only be connected to a properly
grounded wall socket with a protective earth conductor.
To disconnect the mains supply from the centrifuge in the event of malfunction, an
emergency switch separate from the centrifuge must be available. This switch should
be outside the room, where the centrifuge is installed, or next to the entrance to the
room.

2.5 OVERALL VIEW

Main parts of the centrifuge are designated on the following pictures.
CF-220 Centrifuge:
Lid
Main Switch
Window
Control Panel
Auxiliary
Lid Release

Advertisement

Table of Contents
loading