Maintenance Schedule; Malfunctions - Sartorius 16274 Operating Instructions Manual

Stainless steel in-line filter holder
Table of Contents

Advertisement

Available languages

Available languages

Maintenance Schedule

7
Maintenance Schedule
Interval
Before each use of the
device
After each use of the device
Before and after each
sterile filtration
Depending on application.
Must be determined by the
operator
8

Malfunctions

Malfunction
The device has a leak.
The valve has a leak.
The valve cannot be turned.
The membrane filter is torn during
autoclaving.
22
Operating Instructions Stainless Steel In-Line Filter Holder Type 16274
Component
Action
Seals
Check for damage and
replace if necessary
Device
Clean
Device
Perform an integrity test
Seals
Replace
Cause
A seal is damaged.
The seal is damaged.
The valve is dirty.
The membrane filter was not
inserted without tension.
Chapter,
page
5.1.2, 17
6.1, 20
Solution
Check and replace the
seals.
Replace the seal.
Clean the valve.
Insert a new membrane
filter without tension.
Target group
User
User
User
User
Chapter,
page
5.1.2, 17
5.2, 18

Advertisement

Chapters

Table of Contents
loading

Table of Contents