PERFORMANCE CONDITIONS
The electrical characteristics listed in Table 1-1 apply
when a
calibrated probe is used
with
a
calibrated oscIllo-
scope system operating within the environmental c0ndi-
tions stated in Table 1-2.
Items listed in the "Performance Requirement" column
are verifiable qualitative or quantitative limits. Items listed
in the "Supplemental Information" column are not verified
in the "Performance Check
Procedure"
(SectIon 3); they
are either explanatory notes, calibration setup descrip-
tions, performance characteristics for
which
no absolute
limits are specified, or characteristics that are impractical
to check.
The probe's physical characteristics are listed in Table
1-3.
Table 1-1
Characteristic
Performance Requirement
SuppIementaIlnfonnation
Attenuation
lOX
±
1%
at de.
Probe
attached to
11000 series
(system)
plug-in.
Input Resistance
4SOO
±
1%
at de.-
Probe
attached
to
11000 series
(system)
plug-In.
Input Capacitance
Less than
1.6
pF.-
Typically 1.3
pF.
See figure 1-1 for
(1
kHz to
1.5
GHz)
a graph of input Impedance
versus
frequency.
1-2
Specification - P8231
REV AUG 1986