Procedure Steps; Dc Offset Check - Tektronix P6231 Instruction Manual

Hide thumbs Also See for P6231:
Table of Contents

Advertisement

Table 3-1 (cont)
Test Equipment Required
Item Number
Minimum
Purpose
Example of Suitable
and Description
Specification
Test Equipment
8. Adapter
Connectors: GR-to-BNC
HF checks.
Tektronix
Part
Number
female.
017 -0063-00.
9.
Adapter
Connectors: SMA
HF checks.
Tektronix
Part
Number
male-to-GR.
015-1007-00.
PROCEDURE STEPS
b.
Set
the oscilloscope controls
as follows:
1. DC Offset Check
Equipment Required (see Table 3-1):
Oscilloscope (Item
1)
VoltS/Division
Input
Coupling
Time/Division
Trigger Source
Trigger Mode
2V
GND
1
ms
Internal
Auto
Digital Muttimeter
{Item 2)
DC Voltage Source (Item 3)
c.
Vertically
position
the trace to
the
center horizontal
graticule line.
a.
Connect
the probe to an 11 A32 plug-in in an 11000
Series oscilloscope mainframe.
d. Set
the
oscilloscope input coupling to DC.
Performance Cheek Procedure - P6231
3-3