Page 3
P6248 1.7 GHz (Typical) Differential Probe Service Manual Warning The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to all safety summaries prior to performing service.
Page 5
Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
Do Not Operate in an Explosive Atmosphere. Keep Product Surfaces Clean and Dry. Provide Proper Ventilation. Refer to the manual’s installation instructions for details on installing the product so it has proper ventilation. P6248 1.7 GHz Differential Probe Service Manual...
Page 10
DANGER indicates an injury hazard immediately accessible as you read the marking. WARNING indicates an injury hazard not immediately accessible as you read the marking. CAUTION indicates a hazard to property including the product. The following symbol(s) may appear on the product: P6248 1.7 GHz Differential Probe Service Manual...
Use Care When Servicing With Power On. Dangerous voltages or currents may exist in this product. Disconnect power, remove battery (if applicable), and disconnect test leads before removing protective panels, soldering, or replacing components. To avoid electric shock, do not touch exposed connections. P6248 1.7 GHz Differential Probe Service Manual...
Page 12
Service Safety Summary P6248 1.7 GHz Differential Probe Service Manual...
Preface This manual contains specifications and service information for the P6248 differential probe. Related Manuals For operating information, refer to the P6248 Instructions (071-0566-XX). P6248 1.7 GHz Differential Probe Service Manual...
Page 14
Preface viii P6248 1.7 GHz Differential Probe Service Manual...
Specifications The specifications in the following tables apply to a P6248 differential probe installed on a Tektronix TDS694C oscilloscope. When the probe is used with another oscilloscope, the oscilloscope must have an input impedance of 50 Ω. The probe must have a warm-up period of at least 20 minutes and be in an environment that does not exceed the limits described in the environmental specifications...
≤ ±0.5 dB to 850 MHz (deviation from a nominal slope) Bandwidth (probe only) DC to ≥1.7 GHz (- 3dB) @ ≤27 ºC (÷1 range) DC to ≥1.85 GHz (- 3dB) @ ≤27 ºC (÷10 range) <265 ps Rise time (probe and oscilloscope) P6248 1.7 GHz Differential Probe Service Manual...
Performance Verification Use the following procedures to verify the warranted specifications of the P6248 Differential Probe. Before beginning these procedures, photocopy the test record and use it to record the performance test results. The recommended calibration interval is one year.
6. Allow the probe and test equipment to warm up for 20 minutes at an ambient temperature of 20 ºC to 30 ºC (68 ºF to 86 ºF). Perform the verification procedures in order. P6248 1.7 GHz Differential Probe Service Manual...
3. Verify that the output voltage is ≤ ±10 mV. 4. Change the probe attenuation to ÷1. 5. Verify that the output voltage is ≤ ±10 mV. 6. Keep the output connections for the next procedure. P6248 1.7 GHz Differential Probe Service Manual...
14. Calculate the attenuation as follows: (V 1 - V 2) ÷ (V 1 - V 15. Verify that the attenuation is in the range of 0.98 to 1.02. 16. Keep the output connections for the next procedure. P6248 1.7 GHz Differential Probe Service Manual...
÷ V 7. Verify that the attenuation is in the range of 9.5 to 10.5. 8. Remove the Y-lead adapter from the probe and reverse the connection (black to +, red to -). P6248 1.7 GHz Differential Probe Service Manual...
Page 24
19. Measure and record the output voltage as V 20. Calculate the attenuation as V ÷ V 21. Verify that the attenuation is in the range of 0.95 to 1.05. 22. Remove all connections. P6248 1.7 GHz Differential Probe Service Manual...
) that you measured in step 6 represents the rise time of the test system without the probe. The system rise time is used to calculate the probe rise time (t ) in steps 22 and 31. Figure 6: Test system rise time setup P6248 1.7 GHz Differential Probe Service Manual...
Observe proper polarity; the + side of the probe should be on the center square pin of the probe tip adapter. The test setup should now appear as shown. (See Figure 7.) Figure 7: Rise time setup for test system with probe P6248 1.7 GHz Differential Probe Service Manual...
Page 27
10% and 90% amplitude points on the waveform. Record the rise time for ÷10 as t s+p. 31. Calculate the probe rise time using the following formula: 32. Record the calculated probe rise time for ÷10 on the test record. P6248 1.7 GHz Differential Probe Service Manual...
6. Attach the cable from port 2 to the 1103 channel 1 output connector (conventional BNC). Use a male BNC adapter, if necessary. 7. Normalize the network analyzer to remove the loss through channel 1 of the 1103 power supply. P6248 1.7 GHz Differential Probe Service Manual...
Do not allow kinks, loops or other deformities in the probe cable, and maintain a maximum distance between the probe head and the probe control box. P6248 1.7 GHz Differential Probe Service Manual...
Page 30
48 dB). The calculation has been performed in the ÷10 Measured Feedthrough column. 14. Verify that the feedthrough is less than the values in ÷10 Measured Feedthrough column of the table. (See Table 6.) P6248 1.7 GHz Differential Probe Service Manual...
12. Verify that the channel 1 scale factor is 1 V/div. 13. Set the probe attenuation to ÷1. 14. Verify that the channel 1 scale factor is 100 mV/div. This completes the functional check and the performance verification. P6248 1.7 GHz Differential Probe Service Manual...
Page 32
_____________ 1 GHz -30 dB _____________ _____________ ÷10 setting 1 MHz -65 dB _____________ _____________ 100 MHz -45 dB _____________ _____________ 500 MHz -40 dB _____________ _____________ 1 GHz -38 dB _____________ _____________ P6248 1.7 GHz Differential Probe Service Manual...
, Separate Trigger (Square and sine wave pk-pk or Sync output output) 50 Ω terminator 50 Ω ± 1 Ω Tektronix part number 011-0049-01 (needed only if oscilloscope does not support 50 Ω termination) P6248 1.7 GHz Differential Probe Service Manual...
Adjustment Procedures Adjustment locations The location of the adjustments inside the control box are shown below. Figure 9: Adjustment locations P6248 1.7 GHz Differential Probe Service Manual...
4. Adjust the ÷1 offset adjustment for 0 mV, ±1 mV. 5. Change the probe attenuation to ÷10. 6. Adjust the ÷10 offset adjustment for 0 mV, ±1 mV. 7. Keep the output cable set up for the next step. P6248 1.7 GHz Differential Probe Service Manual...
4. Set the probe to ÷1 attenuation. 5. Adjust the power supply to output about 700 mV. 6. Adjust the ÷1 gain adjustment until the output voltage measures the same as the input within ±5 mV. P6248 1.7 GHz Differential Probe Service Manual...
11. Keep the probe output connections for the next step. Offset (Final) NOTE. The offset and gain adjustments interact. 1. Repeat steps 1 through 6 of the Offset (preliminary) procedure. 2. Remove the output cable termination and input for the next procedure. P6248 1.7 GHz Differential Probe Service Manual...
2. Connect a second BNC cable from the function generator Trigger or Sync Output to the External Trigger or Channel 2 input of the oscilloscope. Figure 12: Setup for DC CMRR and AC CMRR adjustments 3. Set the probe to ÷1 attenuation. P6248 1.7 GHz Differential Probe Service Manual...
Page 39
10. Adjust the ÷10 DC CMRR for minimum amplitude in the flat portions of the displayed waveform. This adjustment does not affect the leading edge transitions. 11. Keep the connections for the next procedure. P6248 1.7 GHz Differential Probe Service Manual...
Repeat the DC CMRR adjustment steps 5 through 12. 10. Remove all connections from the probe. Carefully replace the top cover of the control box and the four retaining screws. This completes the adjustment procedures. P6248 1.7 GHz Differential Probe Service Manual...
(See Figure 13.) 2. To replace the pin, insert the new pin into the socket and press it in against a hard surface. Figure 13: Replacing TEKPROBE interface pins P6248 1.7 GHz Differential Probe Service Manual...
2. To replace the collar, align the smaller group of pins with the smaller of the two holes in the interface collar and align the tabs with the slots. Gently press the two pieces together. (See Figure 14.) Figure 14: Replacing the TEKPROBE collar P6248 1.7 GHz Differential Probe Service Manual...
Options Option 95 is a report of the calibration data. This option must be ordered at the time of purchase. P6248 1.7 GHz Differential Probe Service Manual...
Page 44
Options P6248 1.7 GHz Differential Probe Service Manual...
Replaceable Parts Figure 15: P6248 replaceable parts Table 9: P6248 Probe replaceable parts list Fig. & Tektronix Serial Serial index part number number effective discont’d Qty Name & description 15-1 010-6248-00 PROBE,DIFF,ACT:1.7GHZ DIFFERENTIAL,CERTIFICATE OF TRACEABLE CALIBRATION STANDARD,P6248 211-0001-00 SCREW,MACHINE:2-56 X 0.25,PNH,STL CD PL,POZ...
Replaceable Parts Figure 16: P6248 standard accessories P6248 1.7 GHz Differential Probe Service Manual...
Page 47
Serial index Tektronix part number number effective discont’d Name & description 16-1 020-2702-02 ACCESSORY PKG, P6248 (Includes items 2 through 10) 196-3434-00 Y LEAD ADAPTER 196-3504-XX SOLDER-DOWN LEAD, 1 INCH 196-3505-00 SOLDER-DOWN LEAD, 3 INCH 196-3437-XX GROUND LEAD 016-1781-00 ACCESSORY KIT: TIP SAVER, PKG OF 2...
Replaceable Parts Figure 17: P6248 optional accessories P6248 1.7 GHz Differential Probe Service Manual...
Page 49
ADJUSTMENT TOOL: 5 INCH TOTAL, PLASTIC,W/PLASTIC TIP ONE END METAL TIP OPPOSITE END 011-0049-01 TERMN,COAXIAL:50 OHM,2W,BNC 1103 1103 TEKPROBE POWER SUPPLY 012-0076-00 CA ASSY,RF:COAXIAL,RFD,50 OHM,20 L,BNC, MALE,STR,BOTH ENDS,W/STRAIN RELIEF BOOT BOTH ENDS 679-5027-00 CKT BD SUBASSY:1394 ADAPTER P6248 1.7 GHz Differential Probe Service Manual...
Page 50
Replaceable Parts P6248 1.7 GHz Differential Probe Service Manual...
Tektronix Type 1103 Tektronix Type 1103 BNC-to-Type-N coax Type N-male-to-BNC female Tektronix part number adapters (2) 103-0045-00 Tektronix part number BNC cables (2) 50 Ω, 18 inch 012-0076-00 Tektronix part number BNC-to-probe tip adapter 679-4094-00 P6248 1.7 GHz Differential Probe Service Manual...
Adjust the spectrum analyzer input attenuator to accept the high level of 10 dBm. d. Set the reference level to 10 dBm, the center frequency to 1 MHz, and the vertical scale to 10 dB/div. P6248 1.7 GHz Differential Probe Service Manual...
floor. Verify that the change in the level (relative to the level recorded in step 1 f above) is less than or equal to the ÷1 measured feed through column of the table. (See Table 13 on page 41.) P6248 1.7 GHz Differential Probe Service Manual...
Page 54
Carefully dress the probe cable so that it does not cross itself or pass near the BNC to probe tip adapter. Improperly dressing the probe cable can affect the measurements at 1 GHz and above. 22. Repeat steps 10 through 16. P6248 1.7 GHz Differential Probe Service Manual...
Page 55
≤-60 dBm ≤-65 dBm 100 MHz ≥38 dB ≥25 dB ≤-38 dBm ≤-45 dBm 500 MHz ≥35 dB ≥20 dB ≤-35 dBm ≤-40 dBm 1 GHz ≥30 dB ≥18 dB ≤-30 dBm ≤-38 dBm P6248 1.7 GHz Differential Probe Service Manual...