Connect The External Cables; Back Panel Connectors - Agilent Technologies 8860 Installation And First Startup

Gas chromatograph
Hide thumbs Also See for 8860:
Table of Contents

Advertisement

1
Installing the GC
Step 13 Connect the external cables
In addition to the LAN cable as referenced in
Tablet"
sampler (ALS), connecting signal output to integrators, synchronizing the start and end of a
run between various instruments, sensing conditions external to the GC, and controlling
devices external to the GC.

Back panel connectors

The figure below shows the connectors on the back panel of the GC.
ALS1
ALS2
See also
Sampler connectors
If using an ALS, connect it to the GC using the following connectors:
ALS 1 Optional. An injector or 150-position sample tray.
ALS 2 Optional. An injector or 150-Position sample tray.
TRAY
reader/heater/mixer control, if purchased).
The SIG (analog output) connectors
Optional. Use
REMOTE connector
Provides a port to remotely start and stop other instruments using the APG protocol. A
maximum of 10 instruments can be synchronized using this connector. See
Remote Start/Stop Cable"
42
on page 22, other cables may be installed for control of the GC's automatic liquid
"Cable Diagrams"
on page 67.
Optional. The 150-position sample tray (includes optional barcode
SIG1
SIG2
and
for analog output signals.
on page 60 for more detail.
"Connect GC to LAN, Local Computer, or
SIG1
SIG2
BCD
ELVDS1
ELVDS2
LAN
USB1
EVENT
USB2
REMOTE
Installation and First Startup
"Using the