Agilent Technologies 8860 Installation And First Startup page 29

Gas chromatograph
Hide thumbs Also See for 8860:
Table of Contents

Advertisement

1
Installing the GC
3 Install the regulator onto the compressed gas cylinder main fitting.
4 Purge the air from the regulator by repeating the following procedure 5 times:
The image below illustrates a typical pressure regulator installation. In the example shown
below, which uses an optional shutoff valve, open the shutoff valve and leave open during
purging.
Installation and First Startup
Check the thread type. Some regulators use left-hand thread fittings. For left-handed
threads, the nut will have a groove in it.
a Fully close the regulator knob, then open the main tank valve.
b Turn the regulator knob fully counter-clockwise to open the tank main valve,
pressurizing the primary side of the regulator.
c Turn OFF the tank main valve.
d Slowly turn the regulator knob clockwise to release ("bleed off") the gas pressure.
e Close the regulator knob.
Groove indicates
left-hand thread
29