B
Cabling Diagrams and Remote Start/Stop

External event cable, G1530-60590

6
3
The external event cable has two passive relay contact closures with two 24-volt control
outputs. Devices connected to the passive contact closures must be connected to their own
power sources.
The pin assignments for this cable are listed in
Table 15 External events cable
Connector 1 pin
Signal name
24 volts output
1
24 V output 1
2
24 V output 1
3
Ground
4
Ground
Relay contact closures (normally open)
5
Closure 1
6
Closure 1
7
Closure 2
8
Closure 2
When used for external event control, apply label G1580-87200 to identify the cable for EVENT
use.
72
Apply label
G1580-87200
7
8
5
4
1
2
Maximum rating
150 mA
150 mA
48 V AC/DC, 250 mA
48 V AC/DC, 250 mA
Wire terminations
Table
15.
Connector 2, wire color
Controlled by valve #
Yellow
5
Black
6
Red
White
Orange
7
Green
7
Brown or violet
8
Blue
8