B
Cabling Diagrams and Remote Start/Stop
Agilent analog signal cable, G1530-60570
This cable connects an
0 to 10 volts are provided. Connects both GC signal outputs to Agilent 3395B/3396C
integrators, and the 35900 A/D.
Figure 20.
Remote start/stop cable, general use, 35900-60670
The pin assignments for the remote start/stop cable are listed in
Table 12 Remote start/stop cable connections
Connector 1, 9-pin male
1
2
3
4
5
6
7
8
9
68
Analog out
4
Analog output cable to an Agilent product
1
6
9
5
Connector 1
Connector 2, wire color
Black
White
Red
Green
Brown
Blue
Orange
Yellow
Violet
port to an external data system. Both 0 to 1 volt and
35900-60670
Connector 2
Signal
Digital ground
Prepare (low tone)
Start (low tone)
Start relay (closed during
start)
Start relay (closed during
start)
Open circuit
Ready (high true input)
Stop (low tone)
Open circuit
Table
12.
Installation and First Startup