Agilent Technologies 8860 Installation And First Startup page 35

Gas chromatograph
Hide thumbs Also See for 8860:
Table of Contents

Advertisement

1
Installing the GC
4 Connect the calibration gas cylinder output tubing to the hydrogen sensor module.
5 Install the calibration gas cylinder into the stand and secure using the screw.
6 Turn on the supply pressure at the pressure regulator. (You will adjust it later.)
7 Check for leaks in the fittings using a leak detection fluid. Correct any leaks.
Installation and First Startup
Use the tubing, nuts, and ferrules provided in the hydrogen sensor kit.
See the figure under
step
1.
35