WARNING
Risk of injury or damage
When the device is operated at the extremes of the environmental operating specifications (i.e., maximum temperatures and humidity), and in a single fault condition,
which is a single component or performance malfunction, such as a blocked exhaust vent, blocked air intake, or an internal cooling fan failure, the temperature of the air
coming out of the exhaust vents, located on the bottom left and bottom right sides of the unit can reach temperatures capable of causing a burn injury (see stated
temperature and contact time values in the model specific tables below).
Keep exposed body parts, such as hands and feet, a minimum of 46 inches (1.2 meters) away from the exhaust vents to avoid the risk of burns. Single fault conditions
may result in visual and audible alerts and alarms
NOTE – Under normal and single fault conditions, the concentrator releases warm air out the bottom of the unit (exhaust vents) which may discolor temperature
sensitive flooring surfaces. The concentrator should not be used over flooring that is sensitive to heat staining. The Manufacturer is not responsible for flooring that
becomes discolored.
Operate the unit in a cool, dry area with good ventilation, located on a hard surface, avoid thick rugs or carpeting NEVER block the air intake or exhaust vents Keep
the unit a minimum of 12 inches (30.5 cm) away from any wall, draperies, or any other objects that might prevent the proper flow of air in and out of your oxygen
concentrator. Proper air flow is needed to prevent overheating of the oxygen concentrator. DO NOT place the concentrator near any heat source such as hot air
registers or heaters Overheating of the oxygen concentrator may lead to low oxygen output and a risk of burns
The oxygen concentrator should be located in a well-ventilated area. DO NOT operate the unit in a closed or confined space, such as a closet, bathroom, etc. Avoid
operating the device near smoke pollutants and fumes
Under extreme environmental conditions and a single fault condition occurs, the following device surface temperatures may exceed 106 °F (41 °C). See Table 1 below
for the model specific maximum temperature and safe contact guidance:
Table 1 - 525DS Series
Description
Air coming from exhaust vents located on each side
near the bottom of the unit
Oxygen outlet fitting
Power Switch
LED Indicator panel
Cannula at the outlet
Table 2 - 525KS Series
Description
Air coming from exhaust vents located on each side
near the bottom of the unit
LED Indicator panel
Table 3 - 525PS Series
Description
Air coming from exhaust vents located on each side
near the bottom of the unit
Power Switch
LED Indicator panel
CAUTION
Federal (U.S.A.) law restricts this device to sale by or on the order of a physician.
It is very important to follow your oxygen prescription. Do not increase or decrease the flow of oxygen – consult your physician.
Use of harsh chemicals (including alcohol) is not recommended. If bactericidal cleaning is required, a non-alcohol based product should be used to avoid inadvertent
damage
IMPORTANT
It is recommended that the homecare provider lock the flow control knob to prevent inadvertent adjustment. A flow setting other than prescribed may affect the patient
therapy
Do not service or clean this device while in use with a Patient
Installation of 515LF-607 low output flow meter package will cause the low flow alarm to not work and will prevent the device from meeting the requirements of ISO-
80601-2-69:2014 Section 201 13 2 101
The Device is classified as IP21 which means it is protected against finger access to hazardous parts and protected against vertically falling water drops.
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.
This device contains electrical and/or electronic equipment Follow local governing ordinances and recycling plans regarding disposal of device components
SE-525KS-1
Maximum Temperature
°F
147 0
120 7
119 0
129 5
109 1
Maximum Temperature
°F
169 0
110 5
Maximum Temperature
°F
142 2
106 7
108 5
SAVE THESE INSTRUCTIONS.
525DS Series
°C
Max safe contact time
63 9
Less than 1 minute
49 3
Less than 10 minutes
48 3
Less than 10 minutes
54 2
Less than 1 minute
42 8
Less than 10 minutes
525KS Series
°C
Max safe contact time
76 1
Less than 10 seconds
43 6
Less than 10 minutes
525PS Series
Max safe contact time
°C
61 2
Less than 1 minute
41 5
Less than 10 minutes
42 5
Less than 10 minutes
EN - 5