Menu
Customer
adjustment
Rel. humidity
Temperature
Fig. 44
Customer adjustment menu
Fig. 45
1-point adjustment adjustment (offset)
ADJUSTMENT POINT LOW
Adjustment point NEW
current measurement
lower limit output scale
(or last calibration point LOW)
Fig. 46
2-point adjustment procedure
Reference value
Offset
Current
measured value
Reference
value low
2 point adjustment
Reference value
high
i
Calibration status
Reset to factory calibration
adjusted measured value
current measured value
Reference value
1/2 OUTPUT SCALE
ADJUSTMENT POINT HIGH
Reference value
Humidity and Temperature Sensor for High Humidity and Chemically Polluted Conditions | 48
User Manual HTS801
value
Reference value
value
Current measured
value
upper limit output scale
(or last calibration point HIGH)